Page 2 - Kubota - 8730
P. 2

Capacity: 4x1,000mL/130x15mL

          Compact size: 64cmx78cmx86cm (WxDxH)








           Features and Convenient Functions


          ■ Motorized lid lock
            Lid can be opened or closed with one hand.
            To close the lid
            Push the lid gently. Electrical interlock system will
            automatically locks the lid.
            To open the lid
            Press the open key.
            The lid will open automatically after interlock is
            released.
            Optional foot switch is available.

          ■ Acceleration and Deceleration characteristics
            • Acceleration and deceleration curves can be
            varied in 9 steps to accommodate various samples
            and experimental conditions. Acceleration and
            deceleration speeds can be customized for reliable
            centrifugation of samples that easily get stirred
            up.
            • Density gradient method is available.

          ■ 10 channel memories
            10 memory settings are available, with direct keys
            to store, recall and operate.
                                                                                 Foot switch  (Optional)
          ■ Integrator function                                                  Lid opens automatically when the switch is
            Identical applications of total centrifugal force are                pressed. The newly designed shock
            easily reproducible using the g·sec (integrated                      absorber provides smooth and quick lid
            RCF) function.                                                       opening. Hands-free operation facilitates
                                                                                 experiments.
          ■ Sample temperature maintained at 4˚C with
            ambient temperature of 25˚C (Model 8730 only)

          ■ Rotor’s thermal correction                            Acceleration and deceleration profiles for RS-1950.
             (Model 8730 only)                             rpm
            • By setting the temperature of rotor, thermal   Speed
            control will identify the type and speed of rotor    A9    A4  A3  A1                  Natural deceleration
            and correct the temperature of sample to the
            desired value.                                1,000
            • “Internal chamber temperature” or “Rotor     Na                                            Nd
            temperature” can be selected when setting                                    d9   d4    d3     d1  Time
            desired temperature.

                                                                  Acceleration and deceleration profiles for RS-4000.
          ■ Complies with GMP requirements
                                                           rpm
            Tachometer port can be used to measure the    Speed
            actual speed of rotor.                               A9    A4  A3  A1
                                                          2,000
                                                                                                   Natural deceleration
          ■ Safety comes first                            1,000
            Model 8730 and 8622 conform to International   Na                                            Nd
            Safety Standard IEC 61010-2-020
                                                                                         d9    d4    d3    d1  Time
          ■ 4 x 400mL MAP blood bags can be                9 acceleration and deceleration profiles
            centrifuged when using the RS-4000             •  A9 represents the most rapid acceleration, A1 the slowest acceleration.
                                                           •  d9 represents the most rapid deceleration, d1 the slowest deceleration.
            swinging bucket rotor.                         •  For A8 to A4 settings, the acceleration rate changes at 1,000rpm.
                                                           •  For d8 to d1 settings, the deceleration rate changes at 1,000rpm (RS-1950) and
                                                            2000rpm (RS-4000).
                                                           •  For A1 to A3 settings, the point Na where the acceleration rate is changed can
                                                            be set from 0rpm to the maximum speed of rotor.
                                                           •  For d1 to d3 settings, the point Nd where natural deceleration starts can be set
                                                            from 0 rpm to maximum speed of rotor.
   1   2   3   4