Page 2 - Shimadzu TMA-60Series
P. 2

Can be Used for a Wide Range of Measurements

               ●High accuracy measurement of low expansion materials
                                                                          Supports
               A new high-accuracy low-drift displacement sensor has been adopted,   this  Instrument name: TMA-50
                                                                                       Sample name: Super Invar
               dramatically improving measurement accuracy.             measurement    Sample length: 9.990 [mm]
               Even the low expansion metal Super Invar can be measured.
                                                                                       25  ゚C      150  ゚C
               ●Large changes can be measured over a wide dynamic range
                                                                                            0.13um
               Measurement can be carried out over a wide span of ±5 mm, with high-accuracy    0.10 10 -6 /K
               measurement from very small to large deformation.                Measurement of thermal expansion of low expansion metals
               Samples with large deformation, such as separators in lithium ion batteries, can be
               measured.
                                                                                     TMA
                                                                          Supports   1000 um
                                                                                                    a TD
                                                                                      0
               ●Film shrinkage stress can be measured                       this                      b TD
                                                                        measurement  -1000
               Shrinkage stresses can be measured with high sensitivity and high    -2000
               controllability.                                                     -3000               b MD
                                                                                    -4000          a MD
                                                                                    -5000
                                                                                               100
                                                                                         50   Temp  [゚C]  150
            Compact Design                                                        Measurement of Li ion battery separators
               We have achieved a compact design with a small footprint
               (W367 mm, D624 mm).                                        Supports   TMA                LOAD
                                                                                                        g
                                                                                      um
                                                                                                    3.67 g  5.00
                                                                            this        Shrinkage stress
             Example of System Con guration                             measurement  100 .00             0.00
                                                                                                         -5 .00
                                                                                     -0 .00  Displacement of sample
                                                                                                         -10 .00
                                                                                          50 .00  100 .00  150 .00
                                                                                              Temp [゚C]
                                                                                   Measurement of  lm shrinkage stress
                                  Connection
               TMA-60                                  Workstation Software
                                                       LabSolutions™  TA

                                                   Flow Control Unit
                                                   FC-60A

            For measurement below room temperature
            Cooling Options                                       Optional Accessories

            Cooling furnace LTB-60                                Tension chuck for fiber
            ・ Enables TMA measurement over the temperature range of -150 to 500 °C  Used for tension measurement of
            ・  Uses liquid nitrogen                               samples in fiber form.
            ・  Use after manually pouring liquid nitrogen

                                                                  Sample punch for TMA
                                           The cooling furnace     (P/N: 222-12875-91  4×10 mm)
                                           LTB-60 can be used      (P/N: 222-12875-92  4×15 mm)
                                          in place of the normal   (P/N: 222-12875-93  4×20 mm)
                                            heating furnace
                                                                  Used to cut out film samples for tensile
                                                                  measurement with TMA.

            LabSolutions is a trademark of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.





                                                For Research Use Only. Not for use in diagnostic procedures.
                                                This publication may contain references to products that are not available in your country. Please contact us to check the availability of
                                                these products in your country.
                                                Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation,
                                                its subsidiaries or its affiliates, whether or not they are used with trademark symbol “TM” or “®”.
                                                Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether
                                                or not they are used with trademark symbol “TM” or “®”.
                                                Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
        www.shimadzu.com/an/                    The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice.
                                                Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
                                                               © Shimadzu Corporation, 2022 / First Edition: December 2017, 3655-03206-PDFIT, C161-E006A