Page 11 - Shimadzu CELL PICKER
P. 11
Speci cations Installation examples
CELL PICKERʴAUTO CHANGER CELL PICKER
™
Recommended Microscope Olympus CKX 53
Number of Vessels that Up to 8
can be Registered for Picking Note: The only supported multi-well plates are 6-well plates.
Number of Vessels that Up to 8
can be Registered for Seeding Note: Not dishes however. None
Automated Seeding Yes No
Related Item
Recommended Pipette Tip Manufacturer: QSP (Thermo Fisher Scienti c)
(for 200 µL) Model: TW110-96RNS-Q
Desktop Clean Bench dedicated for
Speci cations
CELL PICKER (PCV-801TP-S3) Size: Outside Diameter
Picking mode: 5/10/15 µL Special clean bench that can be loaded on a W740 mm × D650 mm × H890 mm
Suction amount Inside
Removal mode: 5 µL laboratory desk. W690 mm × D520 mm × H605 mm
Air ow system: Vertical air ow system
ɾ Ideal size for CELL PICKER.
Room cleanliness: ISO Class 4
Weight 15.5 kg (AUTO CHANGER: 7 kg) 8.5kg ɾ Utilize CELL PICKER easily due to adoption of a slide Elements of dust-collecting: Pre- lter + HEPA lter
3
table. Air ow: 6.4 m / min ± 20%
Power Supply: 65 W
ɾ Fun Filter Unit maintains cleanliness when opening Weight: Approximately 75 Kg
Size W 585 mm×D536 mm×H456 mm W 448 mm×D536 mm×H456 mm the front shutter. Note: Slide table, Internal plug × 4
(not including microscope) ʢW 385 mm×D355 mm×H372 mmʣ ʢW 248 mm×D355 mm×H372 mmʣ
100 to 240 V AC, Frequency: 50/60 Hz,
Power Supply
Power consumption: 75 VA
Operating temperature: 10 to 35 °C
Environment
Operating humidity: 20 to 85 %RH
CELL PICKER
10 Automated Picking and Collecting of Cell Colonies 11